Laboratory parameters

Rig I (Main rig):
  • Test medium: water with max. temperature of 80 °C > 176°F•Flow capacity: 100 to 5.200 m3/h > 1,44 m3/s > 1.140k UK-gallons
  • Max. pressure difference: 6.000 mbar > 87 PSI
  • Min. pressure difference: 5 mbar > 0,07 PSI
  • Max. system pressure: 9 bar A. > 130 PSI
  • Length of test section: 46 meters (May get extended on special request up to 75 Meters)
  • Available length for Meter Runs: 39 Meters (May get extended on special request up to 68 Meters)
  • Capacity of etalon weight tank: 35.000 kg → 38.5 tn
  • Size of a meter to be calibrated: DN 80 to DN 1.400 > 3” to 56”
  • Expanded uncertainty of “C” (Coefficient of Discharge)
    • 0,22% - 0,5% … using comparison method
    • 0,09% - 0,15% … using gravimetric method

Rig II (Small rig):
  • Test medium: water with max. temperature of 95 °C > 200°F•Flow capacity: 0,1 to 150 m3/h > 0,04 m3/s > 33k UK-gallons
  • Max. pressure difference: 6.000 mbar > 87 PSI
  • Min. pressure difference: 5 mbar > 0,07 PSI
  • Max. system pressure: 9 bar A. > 130 PSI
  • Length of test section: 8 meters
  • Available length for Meter Runs: 5 Meters
  • Size of a meter to be calibrated: DN 15 to DN 100 > ½” to 4”
  • Expanded uncertainty of “C” (Coefficient of Discharge)
    • 0,22% - 0,5% … using comparison method