Scope of Accreditation

Field of measured value: Flowrate

Calibration temperature: (20 ÷ 80) °C

Measured Quantity: Water Flowrate - Mass

Range: (11,1 ÷ 1.166) kg/s

Calibration & Measurement Capability (CMC) Expressed as an Expanded Uncertainty (k=2) : 0,10 % of measured quantity

Measured Quantity: Water Flowrate - Volume

Range: (40 ÷ 4.245) m3/h

Calibration & Measurement Capability (CMC) Expressed as an Expanded Uncertainty (k=2) : 0,10 % of measured quantity

Range: (1 ÷ 150) m3/h

Calibration & Measurement Capability (CMC) Expressed as an Expanded Uncertainty (k=2) : 0,20 % of measured quantity.

 

Methods identification

Calibration Procedure 01 (EN 14154-1+A2, EN 14154-2+A2, EN 14154-3+A2, EN 29104, EN ISO 6817, EN 24185+AC)

Calibration Procedure 02 EN ISO 5167-1, EN ISO 5167-2, EN ISO 5167-3, EN ISO 5167-4, EN 24185+AC, ASME PTC-6, ASME PTC-19.5